Key Specifications
Adapter Types:
1S09168 Type 2 (M4)
1S09168 Type 2 (B2)
KAVO LED Coupling
NSK LED Coupling
NSK QD Coupling
Working Pressure:
Input Water Pressure: 0.7bar to 2.2bar (70-220Kpa)
Input Air Pressure: 3.5bar to 4.5bar (350-450)
Working Mode:
Continuous/Intermittent Operation
Weight:
About 0.150kg
Connectors:
5 Different Connectors
Operation Environment:
Environment Temperature: 10°C-40°C
Relative Humidity: 30%-75%
Atmospheric Pressure: 50kPa-106kPa
Precautions:
Forbidden for hemophilia patients and those with heart pacemakers.
Caution for doctors with heart pacemakers.
Caution for heart disease patients, pregnant women, and children.
Forbidden for patients with respiratory diseases like asthma and chronic bronchitis.
These specifications highlight the device's dimensions, adapter compatibility, working pressures, weight, connectors, operation environment, and crucial precautions for safe and effective use in dental applications.
Direction To Use
Installation:
- Set up external dental devices, including the water supply and air supply systems.
- Ensure the water pressure is between 0.7bar to 2.2bar and adjust the air supply pressure to 3.5bar to 4.5bar.
- Connect the Dental Air Polisher securely to the handpiece pipeline.
Powder Loading:
- After connecting the Dental Air Polisher, unscrew the powder storage chamber cover.
- Add the recommended amount of powder into the chamber.
- Clear any powder residues at the thread position and tighten the cover.
Product Operation:
- Before treating the patient, spray the oral cavity for 1-3 seconds to ensure even air and water spray.
- Hold the handpiece in a pen-holding position.
- Maintain a distance of 3-5mm between the spray head air outlet and the tooth surface during air polishing.
- Follow recommended angles (30-60 degrees) for air polishing direction.
Maintenance:
- Clear powder residues regularly.
- Open the powder storage chamber cover to clear the remaining powder.
- Blow off powder in the chamber with air.
- Wipe off powder residues on the device's surface with alcohol cotton or a dry towel.
Cleaning and Disinfection:
- Remove the handpiece from the Dental Air Polisher.
- Scrub the handpiece and spray head with a soft brush and rinse with pure water.
- Use a CE-certified detergent for cleaning.
- Follow specific procedures for automatic cleaning and avoid ultrasonic cleaning.
Sterilization:
- Sterilize only the handpiece and spray head using steam sterilization procedures.
- Ensure the sterilization temperature does not exceed 138°C.
Usage Precautions:
- Adhere to precautions regarding patient conditions, such as hemophilia, heart pacemakers, and respiratory diseases.
- Regularly check the handpiece for cracks or deformities.