Key Specifications
Input water pressure: 0.7bar to 2.2bar (70-220kPa)
Input air pressure: 3.5bar to 4.5bar (350-450kPa)
Working mode: continuous operation, Intermittent operation
Weight: About 0.150kg
Connector: Meets the requirements of YY / T 0514
Usage: Supragingival air polishing
Operation environment:
Environment temperature: + 10℃~ +40℃
Relative humidity: 30% ~ 75%
Atmospheric pressure:50kPa ~ 106kPa
Direction To Use
Installation and connection:
- The connection between Dental Air Polisher and handpiece pipeline.
- After drying the Dental Air Polisher interface and the handpiece pipeline-interface with an air gun, directly insert the Dental Air Polisher interface into the handpiece pipeline and then tighten the nut of the pipeline.
Powder Loading:
- After connecting the Dental Air Polisher, unscrew the powder storage chamber cover of the gun, put the powder into the chamber, clear the powder at the thread position of the powder storage chamber cover, and tighten the cover.
Supragingival air-polishing:
- After loading the powder, please spray in the container outside the oral cavity for 1~3 seconds in advance to ensure that the air and water can be sprayed evenly before treating the patient.
- Generally, the handpiece is held in a pen-holding position.
- When using a Dental Air Polisher for normal air polishing, it is recommended that the spray-head air outlet and the tooth surface be kept at a distance of 3-5mm, and the air polishing direction is recommended to be at an angle of 30-60° with the tooth surface.
Note:
- It is forbidden to pull out the tail plug when the handpiece is working.
- The only subgingival powder can be used for subgingival air polishing.
Instructions for main components:
- Handpiece: Working part, the spray head of which can be pulled out by rotation. In case of blockage, the user can rotate the spray head to loosen and pull out it, and then use the needle to negotiate. The spray head can be sterilized at high temperatures.
Powder storage chamber:
- Working part, used for powder storage.
- Nozzle: Disposable accessory.
Maintenance:
Powder clearing
- Open the powder storage chamber cover and clear the remaining
powder on it. - Blow off the powder remaining in the powder storage chamber with air.
- Wipe off the powder residues on the surface of the Dental Air Polisher with alcohol cotton or a dry towel.